Sterilization Guidelines

Introduction
Equipment Classification
Levels of Disinfection
Cleaning/Reprocessing of GI Endoscopes
Glutaraldehyde
Methods of Sterilization and Disinfection


Select Chapter:




Search Manual:




Yale New Haven Hospital
QISS
GB 325
New Haven, CT
06504 USA

Dr. Jeff Topal
688-4634




Methods of Sterilization

Method of Sterilization Concentration or Temperature Activity level
Heat >250 F. (121 C),
pre-vacuum cycle **
Sterility
Moist Heat 271 F (132 C)** Sterility
Dry Heat 171 C x 1 hr.

160 C x 2 hr.

121 C x > 16 hr.

Sterility
Gas 450-500 mg/liter at Sterility
Ethylene Oxide (ETO) 55-60 C ** Sterility
**Time variable. Follow sterilizer manufacturer's recommendations for specific exposure and aerations times.